Label requirements · Spain
Spain Medical Device labeling requirements
Everything a label for medical devices sold in Spain must carry, item by item with the article of law it comes from. Paste your own label below: Drict shows what it found, on which line, and what is missing.
5 mandatory items · 6 only if they apply
Your points to check appear here.
Drict looks for the items each label law names, by key words. It is not legal advice and does not say a label is compliant: read the official text, and ask the authority or a specialist before you print.
Medical devices — what a Spain label must show
| What the label must show | Law |
|---|---|
| Manufacturer’s name and registered address | MDR Annex I 23.2(c) |
| Lot number (LOT) or serial number (SN) | MDR Annex I 23.2(g) |
| UDI carrier | MDR Annex I 23.2(h), Art. 27 |
| Use-by date, or the date of manufacture when there is none | MDR Annex I 23.2(i)(j) |
| The indication that it is a medical device (“MD”) | MDR Annex I 23.2(q) |
Only when it applies
These items are needed only for some products, claims or packs — read each one against your product.
| What the label must show | Law |
|---|---|
| Authorised representative in the EU (“EC REP”), if the manufacturer is outside the Union | MDR Annex I 23.2(d) |
| Importer’s name and address, for an imported device | MDR Art. 13(3) |
| Sterile state and method of sterilisation | MDR Annex I 23.2(l) |
| Special storage or handling conditions | MDR Annex I 23.2(k) |
| Warnings and precautions to bring to the user’s attention | MDR Annex I 23.2(m) |
| Single use (“Do not re-use”) | MDR Annex I 23.2(n) |
Language: No Spanish found — devices sold in Spain are labelled in Spanish (MDR Art. 10(11), Real Decreto 192/2023).
Check by eye: The device’s name and what it is for (23.2(a)(b)); the CE marking, with the notified body’s four-digit number where one was involved (Art. 20); ISO 15223-1 symbols (factory, LOT, SN, hourglass, MD, EC REP, STERILE, do not re-use) are pictures — Drict reads only words, so a symbol used alone shows as not found.
Official sources
Questions
What must a label for medical devices sold in Spain show?
Manufacturer’s name and registered address — MDR Annex I 23.2(c); Lot number (LOT) or serial number (SN) — MDR Annex I 23.2(g); UDI carrier — MDR Annex I 23.2(h), Art. 27; Use-by date, or the date of manufacture when there is none — MDR Annex I 23.2(i)(j); The indication that it is a medical device (“MD”) — MDR Annex I 23.2(q). Items needed only in some cases are listed separately on this page; check the official text before you print.
What language must the label be in for Spain?
No Spanish found — devices sold in Spain are labelled in Spanish (MDR Art. 10(11), Real Decreto 192/2023).
Which law sets the label rules for medical devices in Spain?
The points on this page come from: Regulation (EU) 2017/745 on medical devices (MDR). Each item names its article; the official texts are linked under Official sources.
Does a clean Drict report mean my label is legal in Spain?
No. Drict finds the particulars the law names by their usual words and links the law; it cannot judge claims, figures, symbols or letter sizes. Use it as a first pass before a regulatory review, never instead of it.
Is the check free?
Yes. Paste the label text and check one country at a time, free and without an account; the text never leaves your browser. Drict Pro reads photos and PDFs and checks several countries at once.